Back to signals
FDA device recallVulnerabilitiesPublished October 8, 2025

Canon Medical System, USA, INC.: Interventional Fluoroscopic X-Ray System

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Open the source record
Tags
FDA recall, medical device, class 2, Radiology, Radiation Control for Health and Safety Act, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security; CISM 3: Information Security Program
Source record id
Z-0037-2026
First seen by InfoSec Signals
9/23/2026

Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.