Back to signals
FDA device recallVulnerabilitiesPublished October 21, 2025

Wipro GE Healthcare Private Ltd.: Display, Cathode-Ray Tube, Medical

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Open the source record
Tags
FDA recall, medical device, class 2, Cardiovascular, Software design, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 4: Communication and Network Security; CISSP 7: Security Operations
Source record id
Z-0214-2026
First seen by InfoSec Signals
9/23/2026

Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.