Back to signals
FDA device recallVulnerabilitiesPublished November 5, 2025

Tandem Diabetes Care, Inc.: Alternate Controller Enabled Insulin Infusion Pump

Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

Open the source record
Tags
FDA recall, medical device, class 2, Clinical Chemistry, Software design, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security
Source record id
Z-0427-2026
First seen by InfoSec Signals
9/23/2026

Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.

Tandem Diabetes Care, Inc.: Alternate Controller Enabled Insulin Infusion Pump | InfoSec Signals