FDA device recallVulnerabilitiesPublished December 19, 2025

CareFusion 303, Inc.: Cabinet, Table And Tray, Anesthesia

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

Open the source record
Tags
FDA recall, medical device, class 1, Anesthesiology, Device Design, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security