Securitas Healthcare LLC: System, Environmental Control, Powered

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Open the source record
Tags
FDA recall, medical device, class 2, Physical Medicine, Under Investigation by firm, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 4: Communication and Network Security; CISM 3: Information Security Program