Back to signals
FDA device recallVulnerabilitiesPublished February 2, 2026
Intuitive Surgical, Inc.: System, Surgical, Computer Controlled Instrument
Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
Open the source record- Tags
- FDA recall, medical device, class 2, Gastroenterology, Urology, Process control, Open, Classified
- Related exam domains
- CISSP 3: Security Architecture and Engineering; CISSP 7: Security Operations; CISSP 8: Software Development Security; CISM 3: Information Security Program
- Source record id
- Z-1232-2026
- First seen by InfoSec Signals
- 9/23/2026
Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.