Back to signals
FDA device recallVulnerabilitiesPublished July 8, 2026
Medtronic MiniMed, Inc.: Alternate Controller Enabled Insulin Infusion Pump
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
Open the source record- Tags
- FDA recall, medical device, class 2, Clinical Chemistry, Software design, Open, Classified
- Related exam domains
- CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security; CISM 2: Information Security Risk Management
- Source record id
- Z-2698-2026
- First seen by InfoSec Signals
- 9/23/2026
Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.