PIE Medical Imaging B.V.: Automated Radiological Image Processing Software

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Open the source record
Tags
FDA recall, medical device, class 2, Radiology, Software change control, Open, Classified