Beckman Coulter Mishima K.K.: Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Updated under an hour ago ()
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
- Tags
- FDA recall, medical device, class 1, Clinical Chemistry, Software design, Open, Classified
- Related exam domains
- CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security; CISA 3: Information Systems Acquisition, Development and Implementation; CISA 5: Protection of Information Assets
- Source record id
- Z-3332-2026
- First seen by InfoSec Signals
- 10/9/2026
Source: openFDA device recalls, U.S. Food and Drug Administration about this source more from it