CVE-2026-18164Flow Neuroscience FL-100 Use of Hard-coded Credentials
Updated under an hour ago ()
Does this affect you?
It affects Flow Neuroscience FL-100 and 1 other product. Add what you run to find out whether that includes you.
No exploitation reported and no EPSS estimate yet.
Do this next
Free account, 1 product
An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state.
- CVSS
- 8.1 HIGH (v3.1, CNA)CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
- Weakness (CWE)
- CWE-798
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- Not loaded for this CVE
- Assigned by
- icscert
- NVD analysis
- Deferred
- Record last updated
- Aug 14, 2026
- First seen by InfoSec Signals
- Oct 9, 2026
Affected products
Flow Neuroscience FL-100 flow_neuroscience:fl_100 (from the name) | before July 2026 | CNA record |
Halo Neuroscience FL-100 halo_neuroscience:fl_100 (from the name) | before July 2026 | CNA record |
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-225-01
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-225-01.json
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.