CVE-2025-54304An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices.
Updated under an hour ago ()
Does this affect you?
It affects thermofisher ion_torrent_onetouch_2_firmware. Add what you run to find out whether that includes you.
No exploitation reported and no EPSS estimate yet.
Do this next
Free account, 1 product
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
- CVSS
- 9.8 CRITICAL (v3.1, NVD)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-200
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- Not loaded for this CVE
- Assigned by
- Not given
- NVD analysis
- Analyzed
- Record last updated
- Jun 17, 2026
- First seen by InfoSec Signals
- Oct 9, 2026
Affected products
thermofisher ion_torrent_onetouch_2_firmware thermofisher:ion_torrent_onetouch_2_firmware | all versions | NVD |
References
- https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdfproduct
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdfvendor-advisory
- https://www.thermofisher.com/order/catalog/product/4474779product
The CVE record on cve.org NVD's entry