CVE-2023-31222Medtronic Paceart MSMQ Deserialization of Untrusted Data
Updated under an hour ago ()
Does this affect you?
It affects Medtronic Paceart Optima. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 28.49%, higher than 98% of scored CVEs. That is high: schedule it on your emergency patch path.
Do this next
Free account, 1 product
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.
- CVSS
- 9.8 CRITICAL (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-502
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 28.49% (percentile 98)
- Assigned by
- Medtronic
- NVD analysis
- Modified
- Record last updated
- Nov 26, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Medtronic Paceart Optima medtronic:paceart_optima (from the name) | Versions 1.11 and earlier | CNA record |
medtronic paceart_optima medtronic:paceart_optima | before 1.12 | NVD |
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/paceart-optima-system.htmlmitigation, vendor-advisory
The CVE record on cve.org NVD's entry