CVE-2023-30561Lack of Cryptographic Security of IUI Bus
Updated under an hour ago ()
Does this affect you?
It affects Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.19%, higher than 8% of scored CVEs.
Do this next
Free account, 1 product
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
- CVSS
- 6.1 MEDIUM (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:H
- Weakness (CWE)
- CWE-311
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.19% (percentile 8)
- Assigned by
- BD
- NVD analysis
- Modified
- Record last updated
- Oct 22, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 becton_dickinson_and:bd_alarisa_point_of_care_unit_pcu_model_8015 (from the name) | through 12.1.3 | CNA record |
bd alaris_8015_pcu_firmware bd:alaris_8015_pcu_firmware | through 12.1.3 | NVD |
References
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mxmitigation, vendor-advisory
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.