CVE-2023-1966CVE-2023-1966
Updated under an hour ago ()
Does this affect you?
It affects Illumina iScan Control Software and 21 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.92%, higher than 59% of scored CVEs.
Do this next
Free account, 1 product
Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, which could allow an attacker to change settings, configurations, software, or access sensitive data on the affected product.
- CVSS
- 7.4 HIGH (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
- Weakness (CWE)
- CWE-250
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.92% (percentile 59)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Jan 16, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Illumina iScan Control Software illumina:iscan_control_software (from the name) | 4.0.0; 4.0.5 | CNA record |
Illumina MiniSeq Control Software illumina:miniseq_control_software (from the name) | 2.0 | CNA record |
Illumina iSeq 100 illumina:iseq_100 (from the name) | all versions | CNA record |
Illumina MiSeq Control Software illumina:miseq_control_software (from the name) | 4.0 (RUO Mode) | CNA record |
Illumina MiSeqDx Operating Software illumina:miseqdx_operating_software (from the name) | 4.0.1 | CNA record |
Illumina NextSeq 500/550 Control Software illumina:nextseq_500_550_control_software (from the name) | 4.0 | CNA record |
Illumina NextSeq 550Dx Control Software illumina:nextseq_550dx_control_software (from the name) | 4.0 (RUO Mode) | CNA record |
Illumina NextSeq 550Dx Operating Software illumina:nextseq_550dx_operating_software (from the name) | from 1.0.0 through 1.3.1; 1.3.3 | CNA record |
Illumina NextSeq 1000/2000 Control Software illumina:nextseq_1000_2000_control_software (from the name) | through 1.4.1 | CNA record |
Illumina NovaSeq 6000 Control Software illumina:novaseq_6000_control_software (from the name) | through 1.7 | CNA record |
Illumina NovaSeq Control Software illumina:novaseq_control_software (from the name) | 1.8 | CNA record |
illumina iscan_firmware illumina:iscan_firmware | 4.0.0; 4.0.5 | NVD |
illumina iseq_100_firmware illumina:iseq_100_firmware | all versions | NVD |
illumina miniseq_firmware illumina:miniseq_firmware | from 2.0 | NVD |
illumina miseq_firmware illumina:miseq_firmware | from 4.0 | NVD |
illumina miseqdx_firmware illumina:miseqdx_firmware | from 4.0.1; 4.0 | NVD |
illumina nextseq_500_firmware illumina:nextseq_500_firmware | 4.0 | NVD |
illumina nextseq_550_firmware illumina:nextseq_550_firmware | 4.0 | NVD |
illumina nextseq_550dx_firmware illumina:nextseq_550dx_firmware | from 1.0.0 through 1.3.1; from 1.3.3; 4.0 | NVD |
illumina nextseq_1000_firmware illumina:nextseq_1000_firmware | 1.4.1 | NVD |
illumina nextseq_2000_firmware illumina:nextseq_2000_firmware | 1.4.1 | NVD |
illumina novaseq_6000_firmware illumina:novaseq_6000_firmware | through 1.7; 1.8 | NVD |
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01third-party-advisory, us-government-resource
- https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.htmlvendor-advisory
The CVE record on cve.org NVD's entry