CVE-2023-0888Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
Updated under an hour ago ()
Does this affect you?
It affects B. Braun Space Battery Pack SP with Wi-Fi and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.60%, higher than 48% of scored CVEs.
Do this next
Free account, 1 product
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- CVSS
- 4.9 MEDIUM (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
- Weakness (CWE)
- CWE-95
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.60% (percentile 48)
- Assigned by
- B.Braun
- NVD analysis
- Modified
- Record last updated
- Feb 27, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
B. Braun Space Battery Pack SP with Wi-Fi b_braun:space_battery_pack_sp_with_wi_fi (from the name) | through 053L000092; through 054U000092 | CNA record |
bbraun battery-pack_sp_with_wifi_firmware bbraun:battery-pack_sp_with_wifi_firmware | through 053l000092; through 054u000092 | NVD |
References
- https://www.bbraun.com/productsecurityproduct
- https://www.bbraunusa.com/productsecurityproduct
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.