CVE-2022-42732A vulnerability has been identified in syngo Dynamics (All versions < VA40G HF01).
Updated under an hour ago ()
Does this affect you?
It affects Siemens syngo Dynamics and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.67%, higher than 51% of scored CVEs.
Do this next
Free account, 1 product
A vulnerability has been identified in syngo Dynamics (All versions < VA40G HF01). syngo Dynamics application server hosts a web service using an operation with improper read access control that could allow files to be retrieved from any folder accessible to the account assigned to the website’s application pool.
- CVSS
- 7.5 HIGH (v3.1, CISA ADP)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- Weakness (CWE)
- CWE-73
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.67% (percentile 51)
- Assigned by
- siemens
- NVD analysis
- Modified
- Record last updated
- Apr 29, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Siemens syngo Dynamics siemens:syngo_dynamics (from the name) | All versions < VA40G HF01 | CNA record |
siemens syngo_dynamics_cardiovascular_imaging_and_information_system siemens:syngo_dynamics_cardiovascular_imaging_and_information_system | before va40g_hf01 | NVD |
References
- https://www.siemens-healthineers.com/en-us/support-documentation/cybersecurity/shsa-741697vendor-advisory
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.