CVE-2022-38100Contec Health CMS8000
Updated under an hour ago ()
Does this affect you?
It affects Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.98%, higher than 62% of scored CVEs.
Do this next
Free account, 1 product
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
- CVSS
- 7.5 HIGH (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
- Weakness (CWE)
- CWE-400
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.98% (percentile 62)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Apr 16, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor contec_health:cms8000_contec_icu_ccu_vital_signs_patient_monitor (from the name) | all versions | CNA record |
contechealth cms8000_firmware contechealth:cms8000_firmware | all versions | NVD |
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01x-refsource-misc, mitigation, third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.