CVE-2022-36385Contec Health CMS8000
Updated under an hour ago ()
Does this affect you?
It affects Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.39%, higher than 31% of scored CVEs.
Do this next
Free account, 1 product
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
- CVSS
- 6.8 MEDIUM (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-284
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.39% (percentile 31)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Apr 16, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor contec_health:cms8000_contec_icu_ccu_vital_signs_patient_monitor (from the name) | all versions | CNA record |
contechealth cms8000_firmware contechealth:cms8000_firmware | all versions | NVD |
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01x-refsource-misc, mitigation, third-party-advisory, us-government-resource