Medium 4.0HealthcarePublished Jun 2, 2026

CVE-2021-4479Dräger Atlan A350 1.00 <= 1.01 DoS via Medibus Interface

Updated under an hour ago ()

Does this affect you?

It affects Dräger Atlan A350. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.24%, higher than 14% of scored CVEs.

Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

CVSS
4.0 MEDIUM (v3.1, CNA)CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weakness (CWE)
CWE-1286
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.24% (percentile 14)
Assigned by
VulnCheck
NVD analysis
Awaiting Analysis
Record last updated
Jun 10, 2026
First seen by InfoSec Signals
Oct 9, 2026

Affected products

Affected products
Dräger Atlan A350
drager:atlan_a350 (from the name)
from 1.00 through 1.01