CVE-2021-38392Improper Access Control for Boston Scientific Zoom Latitude
Updated under an hour ago ()
Does this affect you?
It affects Boston Scientific ZOOM LATITUDE and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.26%, higher than 16% of scored CVEs.
Do this next
Free account, 1 product
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
- CVSS
- 6.5 MEDIUM (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
- Weakness (CWE)
- CWE-284
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.26% (percentile 16)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Sep 17, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Boston Scientific ZOOM LATITUDE boston_scientific:zoom_latitude (from the name) | Model 3120 | CNA record |
bostonscientific zoom_latitude_pogrammer/recorder/monitor_3120_firmware bostonscientific:zoom_latitude_pogrammer/recorder/monitor_3120_firmware | all versions | NVD |
References
- https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01x-refsource-misc, third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.