CVE-2020-16212Philips Patient Monitoring Devices Exposure of Resource to Wrong Sphere
Updated under an hour ago ()
Does this affect you?
It affects Philips Patient Information Center iX (PICiX) and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.37%, higher than 29% of scored CVEs.
Do this next
Free account, 1 product
In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges.
- CVSS
- 6.8 MEDIUM (v3.1, NVD)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-668
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.37% (percentile 29)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Aug 4, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Philips Patient Information Center iX (PICiX) philips:patient_information_center_ix_picix (from the name) | B.02; C.02; C.03 | CNA record |
philips patient_information_center_ix philips:patient_information_center_ix | b.02; c.02; c.03 | NVD |
References
- https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01third-party-advisory, us-government-resource
- https://www.philips.com/productsecurity
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.