Critical 9.8HealthcarePublished Jun 29, 2020

CVE-2020-12040Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication c…

Updated under an hour ago ()

Does this affect you?

It affects baxter sigma_spectrum_infusion_system_firmware. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.95%, higher than 60% of scored CVEs.

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

CVSS
9.8 CRITICAL (v3.1, NVD)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness (CWE)
CWE-319
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.95% (percentile 60)
Assigned by
icscert
NVD analysis
Modified
Record last updated
Aug 4, 2024
First seen by InfoSec Signals
Oct 12, 2026

Affected products

Affected products
baxter sigma_spectrum_infusion_system_firmware
baxter:sigma_spectrum_infusion_system_firmware
from 6.0 through 6.05; 8.0
NVD

References