CVE-2020-12039Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered…
Updated under an hour ago ()
Does this affect you?
It affects baxter sigma_spectrum_infusion_system_firmware. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.35%, higher than 27% of scored CVEs.
Do this next
Free account, 1 product
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
- CVSS
- 2.4 LOW (v3.1, NVD)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
- Weakness (CWE)
- CWE-259
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.35% (percentile 27)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Aug 4, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
baxter sigma_spectrum_infusion_system_firmware baxter:sigma_spectrum_infusion_system_firmware | from 6.0 through 6.05; 8.0 | NVD |
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04x-refsource-misc, third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry