Back to CVE records
Medium 4.0HealthcarePublished Jun 2, 2026

CVE-2019-25723Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

Updated under an hour ago ()

Does this affect you?

It affects Dräger Perseus A500. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.24%, higher than 14% of scored CVEs.

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

CVSS
4.0 MEDIUM (v3.1, CNA)CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Weakness (CWE)
CWE-1286
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.24% (percentile 14)
Assigned by
VulnCheck
NVD analysis
Awaiting Analysis
Record last updated
Jun 3, 2026
First seen by InfoSec Signals
Oct 9, 2026

Affected products

Affected products
Dräger Perseus A500
drager:perseus_a500 (from the name)
2.00; 2.01; 2.02; 2.03 not affected
CNA record

References

  • vendor-advisory
  • third-party-advisory

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.