CVE-2019-25722Dräger SC Monitoring Devices Hard-coded Credentials and DoS
Updated under an hour ago ()
Does this affect you?
It affects Dräger SC 6002XL and 4 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.19%, higher than 8% of scored CVEs.
Do this next
Free account, 1 product
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
- CVSS
- 7.6 HIGH (v3.1, CNA)CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
- Weakness (CWE)
- CWE-798
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.19% (percentile 8)
- Assigned by
- VulnCheck
- NVD analysis
- Awaiting Analysis
- Record last updated
- Jun 3, 2026
- First seen by InfoSec Signals
- Oct 9, 2026
Affected products
Dräger SC 6002XL drager:sc_6002xl (from the name) | SC 6002XL | CNA record |
Dräger SC6802XL drager:sc6802xl (from the name) | SC6802XL | CNA record |
Dräger SC 7000 drager:sc_7000 (from the name) | SC 7000 | CNA record |
Dräger SC8000 drager:sc8000 (from the name) | SC8000 | CNA record |
Dräger SC90000 XL drager:sc90000_xl (from the name) | SC90000 XL | CNA record |
References
- https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdfvendor-advisory
- https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-hard-coded-credentials-and-dosthird-party-advisory
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.