CVE-2019-25718Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass
Updated under an hour ago ()
Does this affect you?
It affects Dräger Infinity Explorer C700 and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.12%, higher than 2% of scored CVEs.
Do this next
Free account, 1 product
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
- CVSS
- 8.4 HIGH (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-451
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.12% (percentile 2)
- Assigned by
- VulnCheck
- NVD analysis
- Analyzed
- Record last updated
- Jul 15, 2026
- First seen by InfoSec Signals
- Oct 9, 2026
Affected products
Dräger Infinity Explorer C700 drager:infinity_explorer_c700 (from the name) | Infinity Explorer C700 | CNA record |
draeger infinity_explorer_c700_firmware draeger:infinity_explorer_c700_firmware | all versions | NVD |
References
- https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdfvendor-advisory
- https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypassthird-party-advisory, vdb-entry
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.