Medium 6.5HealthcarePublished Nov 25, 2019

CVE-2019-18241In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphe…

Updated under an hour ago ()

Does this affect you?

It affects philips intellibridge_ec40_firmware and 1 other product. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.33%, higher than 24% of scored CVEs.

In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.

CVSS
6.5 MEDIUM (v3.1, NVD)CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Weakness (CWE)
CWE-326
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.33% (percentile 24)
Assigned by
icscert
NVD analysis
Modified
Record last updated
Aug 5, 2024
First seen by InfoSec Signals
Oct 12, 2026

Affected products

Affected products
philips intellibridge_ec40_firmware
philips:intellibridge_ec40_firmware
all versions
NVD
philips intellibridge_ec80_firmware
philips:intellibridge_ec80_firmware
all versions
NVD

References

  • x-refsource-misc, third-party-advisory, us-government-resource

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.