CVE-2019-10964Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control
Updated under an hour ago ()
Does this affect you?
It affects Medtronic MiniMed 508 pump and 29 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.23%, higher than 69% of scored CVEs.
Do this next
Free account, 1 product
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
- CVSS
- 7.1 HIGH (v3.1, CNA)CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
- Weakness (CWE)
- CWE-284
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.23% (percentile 69)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- May 22, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Medtronic MiniMed 508 pump medtronic:minimed_508_pump (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 511 pump medtronic:minimed_paradigm_511_pump (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 512/712 pumps medtronic:minimed_paradigm_512_712_pumps (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 712E pump medtronic:minimed_paradigm_712e_pump (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 515/715 pumps medtronic:minimed_paradigm_515_715_pumps (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 522/722 pumps medtronic:minimed_paradigm_522_722_pumps (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 522K/722K pumps medtronic:minimed_paradigm_522k_722k_pumps (from the name) | all versions | CNA record |
Medtronic MiniMed Paradigm 523/723 pumps medtronic:minimed_paradigm_523_723_pumps (from the name) | through Software Versions 2.4A | CNA record |
Medtronic MiniMed Paradigm 523K/723K pumps medtronic:minimed_paradigm_523k_723k_pumps (from the name) | through Software Versions 2.4A | CNA record |
Medtronic MiniMed Paradigm Veo 554/754 pumps medtronic:minimed_paradigm_veo_554_754_pumps (from the name) | through Software Versions 2.6A | CNA record |
Medtronic MiniMed Paradigm Veo 554CM/754CM pumps medtronic:minimed_paradigm_veo_554cm_754cm_pumps (from the name) | through Software Versions 2.7A | CNA record |
medtronic minimed_508_firmware medtronic:minimed_508_firmware | all versions | NVD |
medtronic minimed_paradigm_511_firmware medtronic:minimed_paradigm_511_firmware | all versions | NVD |
medtronic minimed_paradigm_512_firmware medtronic:minimed_paradigm_512_firmware | all versions | NVD |
medtronic minimed_paradigm_712_firmware medtronic:minimed_paradigm_712_firmware | all versions | NVD |
medtronic minimed_paradigm_712e_firmware medtronic:minimed_paradigm_712e_firmware | all versions | NVD |
medtronic minimed_paradigm_515_firmware medtronic:minimed_paradigm_515_firmware | all versions | NVD |
medtronic minimed_paradigm_715_firmware medtronic:minimed_paradigm_715_firmware | all versions | NVD |
medtronic minimed_paradigm_522_firmware medtronic:minimed_paradigm_522_firmware | all versions | NVD |
medtronic minimed_paradigm_722_firmware medtronic:minimed_paradigm_722_firmware | all versions | NVD |
medtronic minimed_paradigm_522k_firmware medtronic:minimed_paradigm_522k_firmware | all versions | NVD |
medtronic minimed_paradigm_722k_firmware medtronic:minimed_paradigm_722k_firmware | all versions | NVD |
medtronic minimed_paradigm_523_firmware medtronic:minimed_paradigm_523_firmware | through 2.4a | NVD |
medtronic minimed_paradigm_723_firmware medtronic:minimed_paradigm_723_firmware | through 2.4a | NVD |
medtronic minimed_paradigm_523k_firmware medtronic:minimed_paradigm_523k_firmware | through 2.4a | NVD |
medtronic minimed_paradigm_723k_firmware medtronic:minimed_paradigm_723k_firmware | through 2.4a | NVD |
medtronic minimed_paradigm_veo_554_firmware medtronic:minimed_paradigm_veo_554_firmware | through 2.6a | NVD |
medtronic minimed_paradigm_veo_754_firmware medtronic:minimed_paradigm_veo_754_firmware | through 2.6a | NVD |
medtronic minimed_paradigm_veo_554cm_firmware medtronic:minimed_paradigm_veo_554cm_firmware | through 2.7a | NVD |
medtronic minimed_paradigm_veo_754cm_firmware medtronic:minimed_paradigm_veo_754cm_firmware | all versions | NVD |
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-508-paradigm.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-178-01
- http://www.securityfocus.com/bid/108926vdb-entry, third-party-advisory
- https://www.us-cert.gov/ics/advisories/icsma-19-178-01third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.