CVE-2018-8868Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
Updated under an hour ago ()
Does this affect you?
It affects Medtronic 24950 MyCareLink Monitor and 3 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.33%, higher than 25% of scored CVEs.
Do this next
Free account, 1 product
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
- CVSS
- 6.2 MEDIUM (v3.1, CNA)CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
- Weakness (CWE)
- CWE-749
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.33% (percentile 25)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- May 22, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Medtronic 24950 MyCareLink Monitor medtronic:24950_mycarelink_monitor (from the name) | all versions | CNA record |
Medtronic 24952 MyCareLink Monitor medtronic:24952_mycarelink_monitor (from the name) | all versions | CNA record |
medtronic 24950_mycarelink_monitor_firmware medtronic:24950_mycarelink_monitor_firmware | all versions | NVD |
medtronic 24952_mycarelink_monitor_firmware medtronic:24952_mycarelink_monitor_firmware | all versions | NVD |
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.