CVE-2018-8853Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system.
Updated under an hour ago ()
Does this affect you?
It affects Philips Brilliance CT Scanners and 4 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.36%, higher than 28% of scored CVEs.
Do this next
Free account, 1 product
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.
- CVSS
- 8.8 HIGH (v3.0, NVD)CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
- Weakness (CWE)
- CWE-250
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.36% (percentile 28)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Sep 16, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Philips Brilliance CT Scanners philips:brilliance_ct_scanners (from the name) | Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. | CNA record |
philips brilliance_firmware_64 philips:brilliance_firmware_64 | through 2.6.2 | NVD |
philips brilliance_ict_sp_firmware philips:brilliance_ict_sp_firmware | through 3.2.4 | NVD |
philips brilliance_ict_firmware philips:brilliance_ict_firmware | through 4.1.6 | NVD |
philips _brilliance_ct_big_bore_firmware philips:_brilliance_ct_big_bore_firmware | through 2.3.5 | NVD |
References
- https://www.usa.philips.com/healthcare/about/customer-support/product-securityx-refsource-confirm, vendor-advisory
- http://www.securityfocus.com/bid/104088vdb-entry, x-refsource-bid, third-party-advisory
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01x-refsource-misc, third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.