CVE-2018-10626Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity
Updated under an hour ago ()
Does this affect you?
It affects Medtronic 24950 MyCareLink Monitor and 3 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 0.36%, higher than 28% of scored CVEs.
Do this next
Free account, 1 product
Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
- CVSS
- 4.4 MEDIUM (v3.1, CNA)CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N
- Weakness (CWE)
- CWE-345
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 0.36% (percentile 28)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- May 19, 2026
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Medtronic 24950 MyCareLink Monitor medtronic:24950_mycarelink_monitor (from the name) | all versions | CNA record |
Medtronic 24952 MyCareLink Monitor medtronic:24952_mycarelink_monitor (from the name) | all versions | CNA record |
medtronic mycarelink_24952_patient_monitor_firmware medtronic:mycarelink_24952_patient_monitor_firmware | all versions | NVD |
medtronic mycarelink_24950_patient_monitor_firmware medtronic:mycarelink_24950_patient_monitor_firmware | all versions | NVD |
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html
- http://www.securityfocus.com/bid/105042vdb-entry, third-party-advisory
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.