CVE-2018-10596Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Updated under an hour ago ()
Does this affect you?
It affects Medtronic 2090 CareLink Programmer and 2 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.32%, higher than 70% of scored CVEs.
Do this next
Free account, 1 product
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
- CVSS
- 7.1 HIGH (v3.1, CNA)CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-923
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.32% (percentile 70)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- May 22, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Medtronic 2090 CareLink Programmer medtronic:2090_carelink_programmer (from the name) | all versions | CNA record |
Medtronic 29901 Encore Programmer medtronic:29901_encore_programmer (from the name) | all versions | CNA record |
medtronic 2090_carelink_programmer_firmware medtronic:2090_carelink_programmer_firmware | all versions | NVD |
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.