Back to CVE records
Medium 6.5HealthcarePublished Apr 30, 2018

CVE-2017-9657Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible.

Updated under an hour ago ()

Does this affect you?

It affects Philips IntelliVue MX40 Patient Worn Monitor and 1 other product. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.77%, higher than 55% of scored CVEs.

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

CVSS
6.5 MEDIUM (v3.0, NVD)CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Weakness (CWE)
CWE-460
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.77% (percentile 55)
Assigned by
icscert
NVD analysis
Modified
Record last updated
Sep 16, 2024
First seen by InfoSec Signals
Oct 12, 2026

Affected products

Affected products
Philips IntelliVue MX40 Patient Worn Monitor
philips:intellivue_mx40_patient_worn_monitor (from the name)
IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18
CNA record
philips intellivue_mx40_firmware
philips:intellivue_mx40_firmware
before b.06.18
NVD

References

  • x-refsource-confirm, vendor-advisory
  • vdb-entry, x-refsource-bid, third-party-advisory
  • x-refsource-misc, third-party-advisory, us-government-resource

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.