CVE-2017-12724A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
Updated under an hour ago ()
Does this affect you?
It affects smiths-medical medfusion_4000_wireless_syringe_infusion_pump. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.32%, higher than 70% of scored CVEs.
Do this next
Free account, 1 product
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
- CVSS
- 8.1 HIGH (v3.0, NVD)CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- Not given
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.32% (percentile 70)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Aug 5, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
smiths-medical medfusion_4000_wireless_syringe_infusion_pump smiths-medical:medfusion_4000_wireless_syringe_infusion_pump | 1.1; 1.5; 1.6 | NVD |
References
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02Ax-refsource-misc, third-party-advisory, us-government-resource
- http://www.securityfocus.com/bid/100665vdb-entry, x-refsource-bid, third-party-advisory
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.