CVE-2014-5432Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication.
Updated under an hour ago ()
Does this affect you?
It affects Baxter SIGMA Spectrum Infusion System and 1 other product. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 2.57%, higher than 85% of scored CVEs.
Do this next
Free account, 1 product
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
- CVSS
- 9.8 CRITICAL (v3.0, NVD)CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-592
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 2.57% (percentile 85)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Aug 6, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
Affected products
Baxter SIGMA Spectrum Infusion System baxter:sigma_spectrum_infusion_system (from the name) | 6.05 (model 35700BAX) with wireless battery module (WBM) version 16. | CNA record |
baxter sigma_spectrum_infusion_system_firmware baxter:sigma_spectrum_infusion_system_firmware | 6.05 | NVD |
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01x-refsource-misc, mitigation, third-party-advisory, us-government-resource
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.